Welch Allyn Connex® ProBP™ 3400
digital blood pressure device
Directions for use
Zobrazit stránku 0
1 2 3 4 5 6 ... 79 80

Shrnutí obsahu

Strany 1 - Directions for use

Welch Allyn Connex® ProBP™ 3400digital blood pressure deviceDirections for use

Strany 2

4 Introduction Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 3 - Contents

SymbolsDocumentation symbolsWarning: The warning statements in this manual identify conditions or practices that couldlead to illness, injury, or deat

Strany 4

Navigation (Up, Down, Left, Right)Shipping, storing, and environment symbolsFragile; handle with care Relative humiditylimitsTemperature limitsLithium

Strany 5

Class II equipment Available for sale onlyupon order of a physicianor licensed health careproviderSerial number ManufacturerReorder number Call for ma

Strany 6

8 Symbols Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 7 - Initial device setup

General warnings and cautionsWARNING The information in this directions for use is a comprehensive guide tothe operation of ProBP 3400. For best resu

Strany 8

WARNING Welch Allyn is not responsible for the integrity of any mountinginstallation. Welch Allyn recommends that customers contact their BiomedicalE

Strany 9 - Introduction

Caution The device is not heat-resistant. Do not autoclave.Caution Use the device within stated operating temperature ranges. The devicewill not mee

Strany 10

12 General warnings and cautions Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 11 - Button symbols

NIBP warnings and cautionsWARNING ProBP 3400 is not intended to measure blood pressure on neonatalpatients. The AAMI SP10:2002 standard defines neona

Strany 12 - Miscellaneous symbols

© 2011 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication,the purchaser of the product is

Strany 13

WARNING NIBP measurements may be inaccurate in the presence of excessivemotion artifact. Minimize extremity and cuff motion during blood pressure rea

Strany 14

Contents checklistUnpack the ProBP 3400 and any applicable accessories and inspect for missing items. Retain theshipping materials in the event of shi

Strany 15 - General warnings and cautions

16 Contents checklist Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 16

Controls and indicators1. Power on/power off button: controls power to the device.2. Blood Pressure Start/Stop button: initiates a new blood pressure

Strany 17

18 Controls and indicators Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 18

Screen elementsThe liquid crystal display (LCD) may indicate any of the following: systolic blood pressure (mmHg orkPa), diastolic blood pressure (mmH

Strany 19 - NIBP warnings and cautions

20 Screen elements Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 20

Connections1. Mounting connection2. Battery door screw3. Blood pressure hose connection port4. USB/external power cord connection port21

Strany 21 - Contents checklist

22 Connections Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 22

Blood pressure hose and cuffIdentify and have available the ProBP 3400, blood pressure cuff, and the blood pressure hose.1. Inspect the blood pressure

Strany 23 - Controls and indicators

ContentsInitial device setup ... 1Introduction ...

Strany 24

Cuff Size SoftDisposableOne-PieceCuff(5 per pack)Reusable One-Piece Cuff(1 per pack)Circumference(cm)Circumference(in)Small child (size 8) Soft-08 Reu

Strany 25 - Screen elements

Power transformerPower transformer and wall plug/line cord assemblyThe power transformer and wall plug are packaged separately and must be assembled p

Strany 26

Charge the deviceCharge the ProBP 3400 with the Welch Allyn-supplied power supply.To charge the device:1. Insert the USB B connector into the USB/exte

Strany 27 - Connections

SetupBefore first-time use of the ProBP 3400, complete the following steps:Insert the batteryThe battery is shipped separately from the ProBP 3400 dev

Strany 28

Mount the deviceFor mounting instructions, see the accessory's directions for use.For mounting options, see the ProBP 3400 approved accessories i

Strany 29 - Blood pressure hose and cuff

Set time and time format1. The Time format field is highlighted. Press the Up navigation or Down navigationbuttons to access the correct format option

Strany 30

30 Setup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 31 - Power transformer

BatteryWARNING Defective batteries can damage the device. If the battery shows anysigns of damage, leakage, or cracking, it must be replaced immediat

Strany 32 - Charge the device

5. Replace the battery compartment door.6. Secure the screw with the screwdriver.The battery charges when the ProBP 3400 is connected to an external p

Strany 33 - Insert the battery

StartupPress the Power on/power off button to turn the device on or off. Upon each power up, thedisplay lights up and the ProBP 3400 displays the mode

Strany 34 - Initial startup

Battery ... 31Startup ...

Strany 35 - Set time and time format

34 Startup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 36

Standby modeThe standby mode conserves battery power. The display goes black, then enters standby modetwo minutes after a reading or last button press

Strany 37

36 Standby mode Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 38

Blood pressure measurementThe preferred blood pressure measurement site for adults and children is the upper arm. Keep thepatient’s arm relaxed and mo

Strany 39

When complete, the ProBP 3400 device displays systolic, diastolic, pulse rate measurements, and ifenabled, MAP calculation.38 Blood pressure measureme

Strany 40

Settings1. From the Home screen, press the Right navigation button. Settings is highlighted.2. Press the Select button. The Settings menu displays.3.

Strany 41 - Standby mode

The default unit of measure is mmHg.Pressure presetsTo select a pressure preset:1. From the Settings screen, use Up or Down navigation buttons to high

Strany 42

Connect the ProBP 3400 to a target deviceWhen the ProBP 3400 is paired to a target device, it will attempt to automaticallyreconnect each time you tur

Strany 43 - Blood pressure measurement

Select dateTo modify the date shown on the Home screen:1. From the Advanced screen, use Up or Down navigation buttons to highlight Date.2. Press the S

Strany 44

NIBP algorithmTo modify the NIBP algorithm used on the device:1. From the Advanced screen, use Up or Down navigation buttons to highlight NIBPalgorith

Strany 45 - Settings

Appendix ... 69ProBP 3400 approved accessories ...

Strany 46 - Bluetooth wireless technology

44 Settings Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 47 - Advanced settings

Data reviewData is accessed through the Review menu. Data can be identified by number, date, time readingwas taken, diastolic/systolic parameters, MAP

Strany 48 - Select time

46 Data review Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 49 - Language

Maintenance and serviceCalibrate the deviceWelch Allyn recommends annual calibration of the ProBP 3400.To calibrate the ProBP 3400,• Use the Welch All

Strany 50

To remove the battery:1. Remove the screw from the battery compartment door with a Phillips screwdriver.2. Remove the battery compartment door.3. Remo

Strany 51 - Data review

10 percent chlorine bleach/90 percent water solution1. Wipe the device with a clean cloth slightly dampened with a 10 percent bleach and 90 percentwat

Strany 52

50 Maintenance and service Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 53 - Maintenance and service

SpecificationsPhysical specificationsPerformanceThis section describe normal ranges for the ProBP 3400 device.Blood Pressure AccuracyBlood pressure ac

Strany 54 - Clean the device

Electrical specificationsPower requirements:Input: 100-240VAC, 0.18A, 50-60HzOutput: 5VDC, 0.5ADegree of protection: Type BF applied partSafety classi

Strany 55 - Clean the accessories

Product disposalDo not dispose of this product as unsorted municipal waste. Prepare this product for reuse orseparate collection as specified by Direc

Strany 56

vi Contents Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 57 - Specifications

54 Specifications Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 58 - Environmental specifications

Standards and complianceGeneral radio complianceThe Bluetooth Wireless Card must be used in strict accordance with the manufacturer’sinstructions as d

Strany 59 - Product disposal

Welch Allyn is not responsible for any radio or television interference caused by unauthorizedmodification of the devices included with this Welch All

Strany 60

Guidance and manufacturer's declarationEmissions and immunity informationElectromagnetic emissionsThe device is intended for use in the electroma

Strany 61 - Standards and compliance

Electromagnetic immunityThe device is intended for use in the electromagnetic environment specified below. The customer or the userof the device shoul

Strany 62 - European Union

Electromagnetic immunitywhereP is the maximum output powerrating of the transmitter in watts (W) anddis the recommended separation distancein meters (

Strany 63 - Electromagnetic emissions

Recommended separation distances between portable and mobile RFcommunications equipment and the deviceNote 1: At 80 MHz and 800 MHz, the separation di

Strany 64 - Electromagnetic immunity

WarrantyWelch Allyn will warranty the ProBP 3400 device and battery to be free of defects in material andworkmanship and to perform in accordance with

Strany 65

62 Warranty Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 66

TroubleshootingInaccurate blood pressure readingsPossible cause Corrective action and explanationIncorrect cuff sizeUse Welch Allyn approved cuffs onl

Strany 67 - Warranty

Initial device setupBefore using ProBP 3400 for the first time, you must configure the device for use. See Setup foradditional information.1

Strany 68

Possible cause Corrective action and explanationrecommends deflation no faster than 3 mmHgper second.Only use a sphygmomanometer that is calibrated.•

Strany 69 - Troubleshooting

Possible cause Corrective action and explanationDevice is being used in inclimate environmentAmbient temperature is outside of operatingrange. Retry m

Strany 70 - No cuff inflation

Possible cause Corrective action and explanationthe Step algorithm to inflate the cuff to a higherpressure, then attempt to measure the blood pressure

Strany 71 - Cuff deflating too slowly

Bluetooth troubleshootingProBP 3400 and the target device will not pairPossible cause Corrective action and explanationBluetooth radio is not enabled

Strany 72 - Device does not turn on

ProBP 3400 and the target device lost connectionPossible cause Corrective action and explanationUnexpected loss in Bluetooth communicationThe ProBP 34

Strany 73 - Bluetooth troubleshooting

AppendixProBP 3400 approved accessoriesFlexiport® cuffs (Latex free)Part Number Model DescriptionReuse-08 Reusable Cuff, reuse, SM CHILD, 2-tubeReuse-

Strany 74

Blood pressure accessories (Latex free)Part Number Model Description3400-30 ProBP 3400 Double tube blood pressure hose (5 ft)3400-31 ProBP 3400 Double

Strany 75 - Appendix

Part Number DescriptionPWCD-5WT-6 Power cord assembly 6, for mobile stand mounted ProBP3400, Australia/New Zealand —OrangePWCD-5WW-C Power cord assemb

Strany 76 - Mounting options

Part Number DescriptionStartup Guide407673 Startup Guide ProBP 3400Directions for Use, Mounting Options104388 Directions for use, Mounting kit instruc

Strany 77 - Literature/Documentation

Model Description34XXST-B* Includes rechargeable lithium-ion battery, size 11 and size 12 FlexiPort® cuffs,mobile stand mounted configuration*Replace

Strany 78 - Configuration options

2 Initial device setup Welch Allyn Connex® ProBP™ 3400 digital blood pressure device

Strany 80

IntroductionThis Directions for use manual is a comprehensive guide designed to help you understand thecapabilities and operation of the ProBP 3400 no