Directions for use Chapter 10 Troubleshooting and maintenance 169
Mobile Acuity LT Systems: Adjust localization parameters
If your Acuity System is a mobile (laptop computer) system, you can adjust these system
parameters.
Time zone
Acuity System language
Amended ID
Micropaq monitor language
Printout paper size
EN1060 compliance
Temperature units
Enable or disable the optional parallel printer
Set facility name (appears on printouts)
Set unit name (appears in upper right corner of Acuity system display)
Set printer ISO encoding
Set barcode variables
Set ECG filter
To adjust Mobile Acuity LT localization parameters
1. Hold down the key on your keyboard, move the cursor to the upper right corner
of the screen, and press the right mouse button.
2. Click Quit Application, and then click Confirm.
3. Wait for the login screen to appear.
The screen appears within one to two minutes.
4. Type “iupdate”, and then press Enter.
5. Follow the prompts to make the desired changes.
6. Press q to quit.
7. Take one of these steps:
If the system prompts you to reboot, take these steps:
a. Choose 2 to quit.
b. Reboot the system.
The system reboots and starts the Acuity application.
If the system prompts you to quit, take these steps:
a. Choose 1 to quit.
Wait for the login prompt.
b. Type “acuity”, and then press Enter.
You are logged in to the Acuity application.