Connex® CSK softwareInstallation guideSoftware version 3.0.XCP 100™ and CP 200™ electrocardiographsSpot Vital Signs® LXi™vital signs deviceAT-2 Plus e

Strany 2

6 Overview Welch Allyn Connex CSK softwareCP 100 and CP 200 configuration examplesStandalone configuration, removable media or wiredClient-server conf

Strany 3 - Contents

Installation guide Overview 7Client-server configuration, wirelessCardioPerfect Workstation (client)No connectivity software installedServerInstallati

Strany 4

8 Overview Welch Allyn Connex CSK software

Strany 5 - Overview

29InstallationSpot LXiBefore you install the Welch Allyn Connex CSK software do the following:• Verify that the connectivity settings in the device so

Strany 6 - System requirements

10 Installation Welch Allyn Connex CSK software2. Select CP100/200 & SPOT LXi. This screen appears:3. Select SPOT LXi. This screen appears:4. Choo

Strany 7 - About Connex CSK software

Installation guide Installation 112. Select the Uninstall button.3. Follow the prompts.

Strany 8

12 Installation Welch Allyn Connex CSK softwareCP 100 or CP 200Before you install the Welch Allyn Connex CSK software do the following:• If connecting

Strany 9 - Subnet #2

Installation guide Installation 13If this screen does not appear, open Connectivity Install on the CD.2. Select CP100/200 & SPOT LXi. This screen

Strany 10

14 Installation Welch Allyn Connex CSK software3. Select CP100/200. This screen appears:4. Choose either Server or Client based on your configuration.

Strany 11

Installation guide Installation 152. Select the Uninstall button.3. Follow the prompts.

Strany 12

ii Welch Allyn Connex CSK software© 2009, Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manua

Strany 13 - Installation

16 Installation Welch Allyn Connex CSK softwareCP 20Before you install the Welch Allyn Connex CSK software do the following:• Verify that the connecti

Strany 14

Installation guide Installation 17The following screen appears:6. Go to File > Options. You will see the following options:You will normally not ne

Strany 15 - 3. Follow the prompts

18 Installation Welch Allyn Connex CSK softwareSetting up the CP 20 for CardioPerfect connectivitySwitch on the CP 20 electrocardiograph and press the

Strany 16 - CP 100 or CP 200

Installation guide Installation 19AT-2 PlusBefore you install the Welch Allyn Connex CSK software do the following:• Verify that the connectivity sett

Strany 17

20 Installation Welch Allyn Connex CSK softwareThe following screen appears:6. Go to File > Options. You will see the following options:You will no

Strany 18

321ConfigurationConfiguring the CP 200 patient-match criteria(resolving patients)When a CP 200 test is imported into the CardioPerfect database, the p

Strany 19

22 Configuration Welch Allyn Connex CSK software3. Save the file. 4. Restart the CPSAPI service. (Go to Control Panel > Administrative Tools > S

Strany 20

Installation guide Configuration 23Configuring the CardioPerfect Workstation for non-Welch Allyn devicesCardioPerfect Workstation uses the SCP-ECG plu

Strany 21 - Activity log

24 Configuration Welch Allyn Connex CSK softwareTo configure the plug-in for automatic ECG importingThis procedure configures the CardioPerfect progra

Strany 22

Installation guide Configuration 25Data Catcher configurationThe Data Catcher application acts as a gateway between devices and external information h

Strany 23 - AT-2 Plus

iiiContents1 - Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1About this guide . . . . . . .

Strany 24

26 Configuration Welch Allyn Connex CSK softwareGeneral settings1. Stop button Selecting this stops the device server portion of Data Catcher. All sta

Strany 25 - Configuration

Installation guide Configuration 27Device network settings1. Device transmission timeoutThis value represents the maximum time, in seconds, that a dev

Strany 26 - 3. Save the file

28 Configuration Welch Allyn Connex CSK softwareInformation network settings1. Data transmission timeoutThis value represents the maximum time, in sec

Strany 27

Installation guide Configuration 297. Host response XSLTThe Host Response XSLT field is used to specify the location of an .xsl file that will be used

Strany 28

30 Configuration Welch Allyn Connex CSK softwareUser interface settingsDebug logsThe Data Catcher application generates a running log with debug infor

Strany 29 - Data Catcher configuration

Installation guide Configuration 31File Importer configurationFile Importer is a software program that allows you to transfer patient vitals from an e

Strany 30 - General settings

32 Configuration Welch Allyn Connex CSK softwareLogs tabThe Logs tab allows you to read logged messages to either confirm a test was imported correctl

Strany 31 - Device network settings

433TroubleshootingAbout the connectivity software modulesThe Connex CSK software comprises a set of modules that are installed separately as needed. T

Strany 32 - Information network settings

34 Troubleshooting Welch Allyn Connex CSK softwareFunctional diagramsSoftware modules in a standalone CardioPerfect Workstation, removable media or wi

Strany 33

Installation guide Troubleshooting 35Software modules in a client-server configuration, removable media or wiredSoftware modules in a client-server co

Strany 34 - Debug logs

iv Contents Welch Allyn Connex CSK software

Strany 35 - File Importer configuration

36 Troubleshooting Welch Allyn Connex CSK softwareSoftware modules in a Spot LXi wireless configurationSoftware modules in a USB configurationDCServer

Strany 36 - Import failures tab

Installation guide Troubleshooting 37Reassigning a server IP addressWho needs this procedure? Users with a client-server configuration wired to the cl

Strany 37 - Troubleshooting

38 Troubleshooting Welch Allyn Connex CSK softwareTroubleshooting chartConditions Causes Actions1. CP 200 View Available Networks screen does not retu

Strany 38 - Functional diagrams

Installation guide Troubleshooting 394. Test send (or search) failed - wireless.Outside range of all wireless networks.Move within the wireless networ

Strany 39

40 Troubleshooting Welch Allyn Connex CSK softwareConditions (continued) Causes (continued) Actions (continued)6. Test send failed - removable mediaCo

Strany 40

11OverviewAbout this guideThis guide describes how to install, configure, and troubleshoot the Welch Allyn Connex CSK (Connectivity Solutions Kit) sof

Strany 41

2 Overview Welch Allyn Connex CSK softwareSystem requirementsProcessorRAMDisc space400 MHz100 MB300 MB (Application: 20 MB; .NET 280 MBOperating syste

Strany 42 - Troubleshooting chart

Installation guide Overview 3About Connex CSK softwareThe Welch Allyn Connex CSK software is a collection of software modules that enable various Welc

Strany 43

4 Overview Welch Allyn Connex CSK softwareSpot LXi configuration examplesClient-server configuration, wirelessClient-server configuration, wiredServer

Strany 44

Installation guide Overview 5Wireless configuration with multiple subnetsServerInstallation menu option:“Server”Subnet #1Workstation (client)Installat