Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details Host
controller
action
problem
persists,
replace the
module.
8 25 Heater indicates off
when turned on.
Module/
Equipment alert
81
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Indicate
module
error
8 26 HTR_Q on and HTRC is
off, but still has voltage.
Module/
Equipment alert
82
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Power
down
8 27 HTR_Q is tri-stated with
HTRC enabled and has
heater power.
Module/
Equipment alert
83
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Power
down
8 28 Turned Q&C on and
heater voltage not high
enough.
Module/
Equipment alert
84
E5.0
Heater control
error. If the
problem
persists,
replace the
module.
Indicate
module
error
8 29 Heater hardware failsafe
should have turned off
but did not.
Module/
Equipment alert
85 E5.2
Heater control
error. If the
problem
persists,
replace the
module.
Power
down
8 30 Probe over 112° F /
43.3° C.
Module/
Equipment alert
21
E0.4 Probe tip over
temperature.
Heater error or
probe
construction
problem.
Assure that
module is
within
ambient temp
limits. If the
problem
Power
down
Komentáře k této Příručce