Zobrazit stránku 38
SureTemp Plus Module OEM Implementation Communications Protocol
Welch Allyn OEM Technologies Confidential 39
(2936) Send Module Software Version
Causes the module to send the host the software version number of the specified module.
Command Code (hex)
2936
Data
Size (bytes) Type Description
1 int8 Module
0: SureTemp Plus module software version
1: Reserved
2: SureTemp Plus algorithm
3: Reserved
4: Reserved
XX - Undefined
Response Data
Size (bytes) Type Description
N ansichar Bytes to specify the software version.
Notes
The software version is returned as an ASCII string and is of varying length (and format)
dependent on the module specified. The length is guaranteed never to be longer then 127
bytes plus a null terminator. If a module is not present then an array of null characters is
returned. This command returns a “Data Range” error if the specified module is undefined.
Zobrazit stránku 38