Propaq
®
CS Specifications
G
eneral
•
Adult/Pediatric/Neonatal factory/custom
patient modes
•
Electrosurgery interference suppression (ESIS):
all channels except Impedance Respiration
•
All channels electrically isolated
•
Communications option
•
Networks to Acuity
®
Central Station
ECG
•
Automatic 3- and 5-lead detection
•
Selectable leads: I, II, III, aV
R
, aV
L
, aV
F
, V
•
Lead fault detection with auto-reconfigure
•
Bandwidth: 0.5 to 40 Hz monitor mode,
0.05 to 40 Hz extended mode (Adult); 0.5 to
120 Hz monitor mode, 0.05 to 120 Hz extended
mode (Pediatric/Neonate)
•
AAMI 6-pin connector
•
Heart rate measurement range:
25 to 300 bpm (display)
•
Heart rate accuracy: ± 3 bpm or 3%
•
Pacer detection and display
•
Defib sync connector
•
Real-time ECG output, delay < 3 msec
•
ESU and defibrillator protected
•
Sweep speed: 12.5, 25, and 50 mm/sec
•
User-selectable size (mV/cm): 4, 2, 1, 0.5, 0.2
•
QRS tone: high/med/low/off
•
4.2 sec waveform display at 25 mm/sec
•
QRS detector range:
Adult or pediatric mode:
0.22 to 5.0 mV (RTI)
Neonatal mode: 0.1 to 5.0 mV
Neonatal width: 40 to 120 msec
Pediatric width: 40 to 120 msec
Adult width: 70 to 120 msec
Noninvasive Blood Pressure (NIBP)
•
Oscillometric method
•
Automatic (intervals between 1 and 60 min)
and manual modes
•
Quick action NIBP Start/Stop button
•
TurboCuf: 5 min of repeated NIBP readings
•
Systolic, diastolic, mean display
•
On-screen manometer
•
Store and view all data (up to 128 readings)
•
Large NIBP numeric display format
•
Standard cuff sizes: Neonate #1- #5, Infant,
Child, Small Adult, Adult, Large Adult, Thigh
•
Cuff overpressure protection
•
Typical measurement time: 30 to 45 secs
•
Smartcuf
®
NIBP technology
NIBP
Patient mode-specific operation:
Neonate mode
•
Initial cuff inflation pressure: 90 mmHg
•
Maximum allowable cuff pressure: 132 mmHg
•
Systolic range: 25 to 120 mmHg
•
Diastolic range: 10 to 105 mmHg
•
Mean range: 10 to 110 mmHg
Pediatric mode
•
Initial cuff inflation pressure: 120 mmHg
•
Maximum allowable cuff pressure: 170 mmHg
•
Systolic range: 30 to 160 mmHg
•
Diastolic range: 15 to 130 mmHg
•
Mean range: 15 to 140 mmHg
Adult mode
•
Initial cuff inflation pressure: 160 mmHg
•
Maximum allowable cuff pressure: 270 mmHg
•
Systolic range: 30 to 260 mmHg
•
Diastolic range: 20 to 235 mmHg
•
Mean range: 20 to 255 mmHg
Temperature
•
Two YSI 400/700 compatible channels
•
Range: 0 to 50 °C
•
Display: T1, T2 and Delta temp
•
Units: °C or °F
•
Accuracy: ± 0.1 °C (10 to 50 °C),
± 0.2 °C (0 to 10 °C)
•
Optional HP side panel has one temp channel
for YSI 400 probes with HP connector
IBP
•
One or two channels
•
Pressure range: -30 to 300 mmHg
•
Pulse rate measurement range: 25 to 250 bpm
•
Display: 5 scales or semiautomatic rescaling
•
Format: Sys/Dia/Mean, user selectable
•
User-selectable labels: P1, P2, ART,
PA, CVP, ICP; UA, UV Neonate mode only
•
Transducer requirements: 5 µV/V/mmHg
•
Zero adjustment: ± 200 mmHg including
transducer offset
•
Numeric pressure accuracy: ± 2 mmHg or 2%
of reading, plus transducer error
•
Standard 6-pin connector, or HP 12-pin on
HP-compatible side panel option
Impedance Pneumography Respiration
•
Two user-selectable leads: RA-LA, RA-LL
•
Automatic cardiovascular artifact (CVA) rejection
•
Respiration rate range:
Adult/Pediatric modes: 0 (apnea), 2 to 150 RPM
Neonate mode: 0 (apnea), 3 to 150 RPM
•
Respiration detection threshold: 140 milliohms
or 2X CVA, whichever is greater
•
Respiration rate accuracy:
±
2 breaths/min or ± 2 %
•
Apnea alarm delay settings:
Adult/Pediatric modes: 6, 10, 15, 20, 25, 30 sec
Neonate mode: 6, 10, 15, 20 sec
•
Sweep speed: 3.13, 6.25, 12.5 mm/sec
Pulse Oximetry (SpO
2
)
Masimo
®
•
Masimo SET
®
sensors
•
Waveform and pulse amplitude meter
•
Pulse tone pitch indicator
•
Saturation range: 1 to 100%
•
Probe accuracy: 25° to 41°
Nellcor
•
Nellcor
®
OxiMax sensors
•
Waveform and pulse amplitude meter
•
Pulse tone pitch indicator
•
Saturation range: 1 to 100%
•
Probe accuracy: 70 to 100%, 28 to 42 °C,
DS-100A: ± 3
•
Pulse rate measurement range: 25-249 bp
m
•
Pulse rate accuracy: ± 3 bpm
No Motion:
Adult/Pediatric: 70 to 100% ± 2 digits
0 to 69% unspecified
Neonatal: 70 to 100% ± 3 digits
0 to 69% unspecified
During Motion:
Adult/Pediatric/Neonatal: 70 to 100% ± 3 digits
0 to 69% unspecified
•
Pulse rate measurement range
(Adult/Pediatric/Neonatal):
No Motion: 26 to 239 bpm
During Motion: 26 to 239 bpm
•
Pulse rate accuracy:
No Motion: ± 3 digits
During Motion: ± 5 digits
CO
2
Option Display
•
Screen display: CO
2
waveform and EtCO
2
and
INCO
2
(when in alarm) numerics
•
Waveform scale: (maximum) 0 to 100 mmHg,
0 to 14%, 0 to 14 kPa
•
Numeric range display: EtCO
2
: 0 to 99 mmHg,
0 to 13.2 kPa, 0 to 23.1%; INCO
2
: 8
a
to 25 mmHg,
1.1
a
to 5 kPa, 1.1
a
to 5%
•
Units: mmHg, kPa, %; user-selectable
•
Sweep speed: 3.13, 6.25, 12.5 mm/sec;
user-selectable
•
Response modes:
Fast: 15 sec sampling time period; Normal:
30 sec sampling time period; Slow: 45 sec
sampling time period
•
Gas compensation: Off: CO
2
value = calculated
CO
2
value; O
2
> 50%, No N
2
O: CO
2
value =
calculated CO
2
value x 1.03; N
2
O > 50%: CO
2
value = calculated CO
2
value x 0.952
•
Alarm limit ranges: EtCO
2
: 0 to 99 mmHg,
0 to 13.2 kPa, 0 to 13.2%; INCO
2
: 2 to 25 mmHg,
0.2 to 5 kPa, % (no lower limit)
•
Resolution: 1 mmHg
•
Accuracy:
Mainstream
b
:
0 to 30 mmHg, ± 3 mmHg
31 to 99 mmHg, ± 10% of value
Sidestream
c
: 0 to 30 mmHg, ± 3 mmHg
31 to 99 mmHg, ±
10% of value
•
Altitude Error: ± 0.4%/1,000 ft (304.8 m)
Breath Rate Display
•
Screen display: Numeric
•
Breath rate (BR) source: When CO
2
is active,
CO
2
is the BR source. Otherwise, RESP from
ECG is the RR source
•
Units: Breaths/Minute
•
Range:
Adult/Ped: 0 (apnea), 2 to 150 breaths/min
Neonate: 0 (apnea), 3 to 150 breaths/min
•
Resolution: ± 1 breath/min
•
Accuracy: ± 1 breath/min or ± 5%, whichever
is greater
d
•
Alarm limits range:
Adult/Ped: 2 to 150 breaths/min
Neonate: 3 to 150 breaths/min
CO
2
Performance
•
Per ISO 9918:1993(E)/EN 864:1996
Apnea Alarms and Tickets
•
Apnea ticket: Set to auto print after apnea event
and after 1 minute continued apnea
•
Apnea alarm accuracy: ± 2 sec
•
Apnea delay setting:
Adult/Ped: 6, 10, 15, 20, 25, 30 seconds
Neonate: 6, 10, 15, 20 seconds
Barometric Pressure
•
Pressure compensation: automatic
•
Operating range: -2,000 to 15,000 ft,
-610 to 4,572 m, (817 to 429 mmHg)
•
Screen display: Numeric (CO
2
Status Window)
•
Units: mmHg or kPa or %
•
Accuracy: ± 3 mmHg or 2.5% of difference from
calibration pressure, whichever is greater
Sidestream CO
2
Option (SSCO
2
)
•
Sensor type: Sidestream, internal
•
Principle of operation: Nondispersive, infrared,
single-beam, single-path/wavelength, ratiometric
•
Operating ambient temperature: 5 to 40 °C
•
Startup time: 30 seconds typical, 3 minutes
maximum
•
Rise time: 240 ms (10% to 90%) at 175 ml/min
•
Delay time: 1.12 seconds maximum
e
•
Total system response time: 1.36 seconds
(Rise Time and Delay Time)
e
•
Calibration: Verify semiannually, calibrate only
as required
•
Sampling chamber: Internal (replaceable by
service technician)
•
Pneumatic and exhaust system: Integral
•
Barometric pressure compensation: Automatic
•
BTPS, ATPS, STPD: CO
2
value = calculated CO
2
value x 0.977
•
Sampling line: 7-foot sampling line, ID 0.055 in
(1.4 mm), for use with disposable single-use
cannula (CO
2
only or CO
2
sampling/O
2
delivery)
•
Flow rate: 90 or 175 ml/min, user-selectable
a
Lower if in alarm
b
Based on these airway conditions: sensor temperature 42 °C,
airway adapter temperature = 33
°
C, water vapor pressure =
38 mmHg; standard gas mixture = CO
2
in balance air, fully
hydrated at 33
°
C; barometric pressure = 760 mmHg and flow =
60 ml/min
c
Based on the following additional airway conditions: sample line =
7 ft, 0.055 in ID (2.13 m, 1.4 mm ID); sample flow rate =
175 ml/min;
respiratory rate
50 breaths/min, stable to
±
3 breaths/min; inspired/expired time ratio = 1:2;
barometric pressure = 760 mmHg
d
For Sidestream CO
2
, this applies only for BR 60
e
Based on the following additional airway conditions:
sample line = 7 ft, 0.055 in ID (2.13 m, 1.4 mm ID);
sample flow rate = 175 ml/min
Mainstream CO
2
Option (MSCO
2
)
•
CO
2
waveform, EtCO
2
, INCO
2
(INCO
2
when
in alarm), Apnea and Breath Rate display
•
Waveform rise time: < 120 msec (to 90%
after step change)
Left Side AAMI Panel
(HP also available)
Right Side Panel
Mainstream CO
2
Option
Nurse Call Option
Sidestream CO
2
Option
SpO
2
Options
MC 3146 SM2517RevC Propaq CS 2/1/06 9:37 AM Page 1