Propaq
®
CSSpecifications
General
Adult/Paediatric/Neonatal factory/custom
patient modes
Electrosurgery interference suppression (ESIS): all
channels except Impedance Respiration
All channels electrically isolated
Communications option
Networks to Acuity
®
Central Station
ECG
Automatic 3- and 5-lead detection
Selectableleads: I, II, III, aVR, aVL aVF, V
Lead fault detection with auto-reconfigure
Bandwidth: 0.5 to 40 Hz monitor mode, 0.05 to
40 Hz extended mode (Adult); 0.5 to 120 Hz monitor
mode, 0.05 to 120 Hz extended mode (Paediatric/
Neonate)
AAMI 6-pin connector
Heartratemeasurementrange: 25 to 300 bpm
(display)

Heartrateaccuracy: ± 3 bpm or 3%
Pacer detection and display
Defib sync connector
Real-time ECG output, delay < 3 msec
ESU and defibrillator protected
Sweepspeed: 12.5, 25, and 50 mm/sec
User-selectablesize(mV/cm): 4, 2, 1, 0.5, 0.2

QRStone: high/med/low/off
4.2 sec waveform display at 25 mm/sec
QRSdetectorrange:
Adultorpaediatricmode: 0.22 to 5.0 mV (RTI)
Neonatalmode: 0.1 to 5.0 mV
Neonatalwidth: 40 to 120 msec
Paediatricwidth:40 to 120 msec
Adultwidth: 70 to 120 msec
Noninvasive Blood Pressure (NIBP)
Oscillometric method
Automatic (intervals between 1 and 60 min)
and manual modes
Quick action NIBP Start/Stop button
TurboCuf: 5 min of repeated NIBP readings
Systolic, diastolic, mean display
On-screen manometer
Store and view all data (up to 128 readings)
Large NIBP numeric display format
Standardcuffsizes: Neonate #1- #5, Infant,
Child, Small Adult, Adult, Large Adult, Thigh
Cuff overpressure protection
Typicalmeasurementtime:30 to 45 secs
Smartcuf
®
NIBP technology
NIBP
Patient mode-specific operation:
Neonate mode
Initialcuffinflationpressure: 90 mmHg

Maximumallowablecuffpressure:132 mmHg

Systolicrange: 25 to 120 mmHg

Diastolicrange: 10 to 105 mmHg

Meanrange: 10 to 110 mmHg
Paediatric mode
Initialcuffinflationpressure: 120 mmHg

Maximumallowablecuffpressure: 170 mmHg
Systolicrange:30 to 160 mmHg
Diastolicrange: 15 to 130 mmHg
Meanrange: 15 to 140 mmHg
Adult mode

Initialcuffinflationpressure: 160 mmHg
Maximumallowablecuffpressure: 270 mmHg
Systolicrange: 30 to 260 mmHg
Diastolicrange: 20 to 235 mmHg
Meanrange: 20 to 255 mmHg
Temperature
Two YSI 400/700 compatible channels

Range: 0 to 50 °C
Display:T1, T2 and Delta temp
Units:°C or °F

Accuracy: ± 0.1 °C (10 to 50 °C),
± 0.2 °C (0 to 10 °C)
Optional HP side panel has one temp channel for
YSI 400 probes with HP connector
IBP
One or two channels
Pressurerange: -30 to 300 mmHg
Pulseratemeasurementrange: 25 to 250 bpm
Display: 5 scales or semiautomatic rescaling

Format: Sys/Dia/Mean, user selectable
User-selectablelabels:P1, P2, ART,
PA, CVP, ICP; UA, UV Neonate mode only

Transducerrequirements: 5 µV/V/mmHg

Zeroadjustment: ± 200 mmHg including
transducer offset
Numericpressureaccuracy: ± 2 mmHg or 2%
of reading, plus transducer error
Standard 6-pin connector, or HP 12-pin on
HP-compatible side panel option
Impedance Pneumography Respiration

Twouser-selectableleads: RA-LA, RA-LL
Automatic cardiovascular artifact (CVA) rejection
Respirationraterange:
Adult/Paediatricmodes:0 (apnea), 2 to 150 RPM
Neonatemode:0 (apnea), 3 to 150 rpm
Respirationdetectionthreshold:
140 milliohms or 2X CVA, whichever is greater
Respirationrateaccuracy:
± 2 breaths/min or ± 2%
Apneaalarmdelaysettings:
Adult/Paediatricmodes: 6, 10, 15, 20, 25, 30 sec
Neonatemode: 6, 10, 15, 20 sec
Sweepspeed: 3.13, 6.25, 12.5 mm/sec
Pulse Oximetry (SpO
2
)
Masimo
®
Masimo SET
®
sensors
Waveform and pulse amplitude meter
Pulse tone pitch indicator

Saturationrange: 1 to 100%
Probeaccuracy:25° to 41°
NoMotion:
Adult/Paediatric: 70 to 100% ± 2 digits,
0 to 69% unspecified
Neonatal:70 to 100% ± 3 digits,
0 to 69% unspecified
DuringMotion:
Adult/Paediatric/Neonatal:70 to 100% ± 3 digits,
0 to 69% unspecified

Pulseratemeasurementrange
(AdultPaediatric/Neonatal):
NoMotion: 26 to 239 bpm
DuringMotion: 26 to 239 bpm
Pulserateaccuracy:
NoMotion: ± 3 digits
DuringMotion: ± 5 digits
Left Side AAMI Panel
(HP also available)
!
Right Side Panel
Mainstream CO
2
Option
Nurse Call Option
Sidestream CO
2
Option
SpO
2
Options