Welch Allyn PROPAQ CS Uživatelský manuál Strana 4

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 4
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 3
WelchAllynU.K.Ltd.
Cublington Road, Aston Abbotts, Buckinghamshire HP22 4ND, England
Tel: 0 207 365 6780 Fax: 0 207 365 9694
WelchAllynEuropeanCustomerService
Navan Business Park, Dublin Road, Navan, County Meath, Ireland
Tel: +353 46 906 7790 Fax: +353 46 906 7754
WelchAllynPTY.LTD,S.Africa
394 Surrey Ave, Ferndale, Randburg 2194 South Africa
P O Box 809, Fontainebleau, South Africa 2032
Tel: 0 11 777 7555 Fax: 0 11 777 7556
www.welchallyn.com
© 2009 Welch Allyn SM2517EU Rev D
All rights reserved. Acuity, Propaq and Smartcuf are registered trademarks and FlexNet is a trademark of Welch Allyn.
Other trademarks mentioned are for identification purposes only and are owned by their respective companies.
Welch Allyn Acuity
®
Central Monitoring
Station utilizes Aruba Mobile Edge
Architecture for best-in-class HIPAA security,
a superior voice and data network solution
and high availability for critical applications.
Only Masimo SET SpO
2
sensors, including
LNOP durable adhesive and NR value disposable
sensors, should be used with Masimo pulse oxi-
metry. Masimo, Masimo SET, LNOP, and NR are
registered trademarks of Masimo Corporation.
Only Nellcor
®
OxiMax
®
pulse oximetry sensors
should be used with the Nellcor pulse oximetry
option. NELLCOR and OxiMax are registered
trademarks of Nellcor Puritan Bennett, Inc.
All Propaq CS vital signs monitors come standard
with Smartcuf technology, a new standard in
NIBP accuracy. This patented software offers
industry-leading performance and accuracy in
the presence of motion artifacts.
Alarms
Allparameters: upper/lower limits
Allparameters:Adult/Paediatric/Neonate patient
mode-specific limits
Factory default or programmable settings for all
patient modes
Alarmindicator: red
Alarm(s)offindicator:yellow
Audiblealarmtone: high/med/low
Alarmsuspend:90 seconds
One-button Stat Set for all alarm limits
(except Apnea Delay)
EQAlert: yellow
Trends
Tabular numeric format
All parameters trended/viewable
Resolution:non-NIBP trends entered every 2 min
NIBP trends entered after each reading
Duration: 5 hours non-NIBP trends
(up to 150 readings)
8 hours NIBP trends
(up to 128 readings)
Page up/down trend view
Inservice Mode
Includes two sets of simulated patient data including
waveforms, for training and education
Colour Active Matrix Display
Type: TFT (Thin Film Transistor) LCD module
Resolution: 640 x 480 pixels, 1 pixel = R+G+B dots
Activeviewingarea: 170.9 x 129.6 mm
Pixelpitch: 0.267 mm
Viewingangle: U/D 40°, R/L 60° (typical), 10:1
contrast ratio
Contrastratio:150:1 (typical); measured in dark
room at centre of screen
Displaycolour:18-bit (6 bits per primary color)
Waveformsandmatchingnumerics: Predefined

Luminance: 200 cd/m
2
(typical); measured at
saturation point
Responsetime: 40 ms (maximum); “white to black”
Environmental
Operatingtemperature:5 to 40 °C (monitor),
5 to 40 °C (printer)
Operatingrelativehumidity: 15% to 95%,
noncondensing (MIL STD 810E), printer 35%
to 85%, noncondensing (MIL STD 810E)
Operatingaltitude:-610 to 4,572 m
(817 to 429 mmHg)

Shipping/storagetemperature: -20 to 60 °C

Shock: 50 g (monitor); 30 g (expansion module/
printer)

Randomvibration:Designed to meet RTCA
DO-160D, category C (monitor)

Electromagneticcompatibility(EMC):
per EN60601-1-2, 1993
Drip-proof per IEC 529, level IPX1 (monitor)
Physical

Monitor(includinghandle):
Size: 20.8 cm (H) x 24.4 cm (W) x 14.1 cm (D)
Weight:3.4 kg

MonitorwithSpO
2
Module:
Size:20.8 cm (H) x 24.4 cm (W) x 19.7 cm (D)
Weight:4.9 kg

MonitorwithExpansionModule:
(with Printer/SpO
2
/MSCO
2
)
Size:28.8 cm (H) x 24.4 cm (W) x 19.7 cm (D)
Weight:6.5 kg
Power
Rechargeable sealed lead acid internal battery pack
Internal recharger
Inputvoltage:12 to 28 VDC, 25 W
ACpoweradapter: 100 to 120 VAC, 50 to 60 Hz
(North America/Japan); 220 to 240 VAC, 50 to
60 Hz (International)
Operatingtimeonbattery,typical*: Monitor
with SpO
2
option: 4 hours; Monitor with Expansion
Module (Printer, SpO
2
/CO
2
): 3 hours;
Monitor only:2 hours

Batteryrechargetime:6 to 8 hours
(when monitor OFF)

Batteryrechargetime:8 to 12 hours
(when monitor ON)
Certifications/Standards

Complies with relevant AAMI, IEC, EN, CSA
and UL standards. CE Marked according to
the European Medical Device Directive. CSA
Certificate of Compliance for use in the U.S.
and Canada
VA Contract V797P-3486k
DSCP Contract SP0200-97-D-8021
* Fully charged, new battery at 25 °C, NIBP and
Snapshot (if applicable) every 15 minutes, all
accessories used.