Propaq
®
CSSpecifications
Nellcor
•
Nellcor
®
OxiMax sensors
•
Waveform and pulse amplitude meter
•
Pulse tone pitch indicator
•
Saturationrange: 1 to 100%
•
Probeaccuracy:70 to 100%, 28 to 42 °C,
DS-100A: ± 3
•
Pulseratemeasurementrange: 25-249 bpm
•
Pulserateaccuracy: ± 3 bpm
CO
2
Option Display
•
Screendisplay: CO
2
waveform and EtCO
2
and
INCO
2
(when in alarm) numerics
•
Waveformscale: (maximum) 0 to 100 mmHg,
0 to 14%, 0 to 14 kPa
•
Numericrangedisplay:EtCO
2
: 0 to 99 mmHg,
0 to 13.2 kPa, 0 to 23.1%; INCO
2
: 8
a
to 25 mmHg,
1.1
a
to 5 kPa, 1.1
a
to 5%
•
Units: mmHg, kPa, %; user-selectable
•
Sweepspeed:3.13, 6.25, 12.5 mm/sec; user-
selectable
•
Responsemodes:
Fast:15 sec sampling time period;
Normal:30 sec sampling time period;
Slow: 45 sec sampling time period
•
Gascompensation: Off: CO
2
value = calculated
CO
2
value; O
2
> 50%, NoN
2
O: CO
2
value =
calculated CO
2
value x 1.03; N
2
O>50%: CO
2
value = calculated CO
2
value x 0.952
•
Alarmlimitranges:EtCO
2
:0 to 99 mmHg,
0 to 13.2 kPa, 0 to 13.2%; INCO
2
: 2 to 25 mmHg,
0.2 to 5 kPa, % (no lower limit)
•
Resolution: 1 mmHg
•
Accuracy:
Mainstream
b
:0 to 30 mmHg, ± 3 mmHg
31 to 99 mmHg, ± 10% of value
Sidestream
c
: 0 to 30 mmHg, ± 3 mmHg
31 to 99 mmHg, ± 10% of value
•
AltitudeError:± 0.4%/304.8 m
Breath Rate Display
•
Screendisplay: Numeric
•
Breathrate(BR)source:When CO
2
is active,
CO
2
is the BR source. Otherwise, RESP from
ECG is the RR source
•
Units: Breaths/Minute
•
Range:
Adult/Paed: 0 (apnea), 2 to 150 breaths/min
Neonate: 0 (apnea), 3 to 150 breaths/min
•
Resolution: ± 1 breath/min
•
Accuracy: ± 1 breath/min or ± 5%, whichever
is greater
d
•
Alarmlimitsrange:
Adult/Paed: 2 to 150 breaths/min
Neonate:3 to 150 breaths/min
CO
2
Performance
•
Per ISO 9918:1993(E)/EN 864:1996
Apnea Alarms and Tickets
•
Apneaticket: Set to auto print after apnea event
and after 1 minute continued apnea
•
Apneaalarmaccuracy: ± 2 sec
•
Apneadelaysetting:
Adult/Paed: 6, 10, 15, 20, 25, 30 seconds
Neonate: 6, 10, 15, 20 seconds
Barometric Pressure
•
Pressurecompensation: automatic
•
Operatingrange: -610 to 4,572 m
(817 to 429 mmHg)
•
Screendisplay: Numeric (CO
2
Status Window)
•
Units: mmHg or kPa or %
•
Accuracy: ± 3 mmHg or 2.5% of difference from
calibration pressure, whichever is greater
Sidestream CO
2
Option (SSCO
2
)
•
Sensortype: Sidestream, internal
•
Principleofoperation:Nondispersive, infrared,
single-beam, single-path/wavelength, ratiometric
•
Operatingambienttemperature: 5 to 40 °C
•
Startuptime: 30 seconds typical, 3 minutes
maximum
•
Risetime: 240 ms (10% to 90%) at 175 ml/min
•
Delaytime:1.12 seconds maximum
e
•
Totalsystemresponsetime: 1.36 seconds
(Rise Time and Delay Time)
e
•
Calibration: Verify semiannually, calibrate only as
required
•
Samplingchamber:Internal (replaceable by
service technician)
•
Pneumaticandexhaustsystem:Integral
•
Barometricpressurecompensation:Automatic
•
BTPS,ATPS,STPD: CO
2
value = calculated CO
2
value x 0.977
•
Samplingline: 2.13 m sampling line, ID 1.4 mm,
for use with disposable single-use cannula (CO
2
only or CO
2
sampling/O
2
delivery)
•
Flowrate: 90 or 175 ml/min, user-selectable
a
Lower if in alarm
b
Based on these airway conditions: sensor temperature 42 °C, airway
adapter temperature = 33 °C, water vapor pressure =
38 mmHg; standard gas mixture = CO
2
in balance air, fully hydrated at
33 °C; barometric pressure = 760 mmHg and flow = 60 ml/min
c
Based on the following additional airway conditions: sample line =
2.13 m, 1.4 mm ID; sample flow rate =175 ml/min; respiratory rate
< 50 breaths/min, stable to ± 3 breaths/min; inspired/expired time
ratio = 1:2; barometric pressure = 760 mmHg
d
For Sidestream CO
2
, this applies only for BR
<
60
e
Based on the following additional airway conditions:
sample line = 2.13 m, 1.4 mm ID;
sample flow rate = 175 ml/min
Mainstream CO
2
Option (MSCO
2
)
•
CO
2
waveform, EtCO
2
, INCO
2
(INCO
2
when
in alarm), Apnea and Breath Rate display
•
Waveform rise time: < 120 msec (to 90% after
step change)
Mainstream CO
2
Sensor
•
Mainstream NDIR single-beam, single path/
wavelength, ratiometric
•
Warm-uptime(CO
2
sensorandmonitor):
45 sec typical, 3 minutes maximum
•
Automatic zeroing
•
No routine calibration
•
Operatingaltitude: -610 to 4,572 m
(817 to 429 mmHg)
•
Dimensions:2.5 cm H x 2.6 cm W x 2.0 cm L
•
Weight: 12 g
•
Cablelength: 3.05 m nominal
CO
2
Airway Adapters
•
Disposable or multi-use
•
Size: 15 mm ID (meet ISO specifications)
•
Singleormulti-useadult/paediatricairway
adapterforpatientswhoweigh
>
5kg:
Added deadspace: < 6 cc
•
Single-uselow-deadspaceairwayadapterfor
patientswhoweigh<5kg:
Added deadspace: < 0.6 cc
Wireless Option
•
FlexNet 802.11a wireless radio deployed on
dedicated 802.11a or enterprise 802.11a/b/g
networks
•
Industry standard IEEE 802.11a/b/g compliant
•
Frequency:2.4–2.5, 5.15–5.25, 5.25–5.35,
5.47–5.725, 5.725–5.85 GHz RF
•
WirelessmonitorsperAccessPoint:
20 (maximum)
Nurse Call Option
•
Maximumswitchcurrent: 1A
•
Maximumswitchvoltage: 30 V AC/DC
•
Isolation:1500 Vrms
•
Alarmrelay: Energised during apnea alarm or
patient alarm
•
Customisedcable: One end is a 4-pin plug,
compatible with the monitor’s nurse call connector;
the other end must be customised to connect to the
local nurse call system
Printer Option
•
Thermal sensitive dot
•
Numericannotation: date, time, all active
parameters, patient name if connected to Acuity
•
Numberofwaveformsprintedsimultaneously:
up to 3
•
Printspeeds: 6.25, 12.5, 25.0 mm/sec
•
Format: 53 mm wide print area on 60 mm wide paper
Operating modes:
Continuous: start/stop real-time printouts with all
active numerics and up to 3 waveforms
Snapshot: 8 sec (32 sec for CO
2
/Resp) of
immediate history for all active numerics and up to
3 waveforms
Alarmprint: 20 seconds total/12 seconds prior to
alarming parameter
Autoprint:Snapshot printout every 15 or 30 min,
or 1, 2 or 4 hours
Trendprint: On demand or Auto Trend every
4 hours. Up to 8 hours in tabular format.
OxyCRGtrendprint: Print-only function (CO
2
or Resp
must be installed): 2 min printout of trended HR/
PR and SpO
2
numerics and compressed CO
2
or Resp
waveforms (CO
2
is priority source), with annotation and
MIN/MAX tabular report
OxyCRGonalarm:Set to print out 60 sec after
HR/PR or SpO
2
alarm or 75 sec after BR/RR or
apnea alarm
Apneaticket:Auto print HR/PR, SpO
2
and elapsed
time after apnea event and after 1 minute of
continued apnea
Cuffticket:Auto print NIBP, HR/PR, SpO
2
, CO
2
and
RR/BR after each NIBP reading